top of page

Distal femur tumor resection: working with a 5 mm margin

Isabelle Têcheur
4 days ago
2 min read

5 mm. That was the safety margin planned around this distal femur tumor, located on the medial epicondyle. To respect it, the resection relied on a 3D-Cut patient-specific cutting guide.

A margin this thin leaves little room for error. The blade has to follow each cutting plane, and stop exactly where the plane stops. Every detail of the guide was designed around that requirement.


Planning the distal femur tumor resection

The resection was planned with five cutting planes, through an anterior and medial approach. The guide sits around the tumor, so that the resected piece can be cut and then extracted in that same direction.


Eight K-wires, two roles

The guide is held by eight 2.0 mm K-wires.

Four of them are placed at the intersection of the cutting planes, marking where one cut ends and the next begins.

The other four stabilize the guide before cutting and play a second role, as mechanical stops: they stop the blade and prevent unwanted cuts into healthy bone. To work as intended, these stops require a saw blade wider than 8 mm. Some planes may also need an additional instrument, such as a chisel, to complete the cut.


One position, and a contact surface the surgeon can reach

A patient-specific guide is only reliable if it fits the bone in a single position. Here, a small wing follows the medial supracondylar line to lock that position.

But a contact surface only helps if the surgeon can reach it. During the design phase, the question was asked directly: will you be able to reach all the contact surfaces through your approach, especially the wing? The answer was yes, and the guide was validated before manufacturing.


A removable stiffening cylinder

A cylinder connects both ends of the guide to make it more rigid during cutting. It can be removed if needed.


Relying on the plan

With a 5 mm margin, accuracy comes down to details: K-wires that double as mechanical stops, a contact surface checked against the surgical approach, a guide stiffened for the cut. Together, they make it possible to rely on the plan in the operating room.

Planning a bone tumor resection with tight margins? Contact us to discuss your case and see how 3D-Cut can support your planning.

 
 
 

Comments


bottom of page